International Journal of Radiation Research، جلد ۵، شماره ۴، صفحات ۱۸۱-۱۸۶
|
|
|
عنوان فارسی |
|
|
چکیده فارسی مقاله |
|
|
کلیدواژههای فارسی مقاله |
|
|
عنوان انگلیسی |
Radial dose functions of GZP6 intracavitary brachytherapy 60Co sources: treatment planning system versus Monte Carlo calculations |
|
چکیده انگلیسی مقاله |
Background: The Monte Carlo (MC) method is not only used for dose calculations around brachytherapy sources but also for benchmarking treatment planning systems (TPS) calculations. Materials and Methods: Three 60Co sources of GZP6 brachytherapy unit were simulated using MCNP4C MC Code. The radial dose functions were calculated by MC method and GZP6 TPS and were compared. Results: There was a good agreement between MC and TPS calculations for all sources. Discrepancies up to 10% were observed for points close to the sources, but for those farther than 7mm from source center, differences were less than 2% for all sources. Conclusion: Our results showed that GZP6 TPS calculations can accurately be used for dose calculations in brachytherapy treatments for points farther than 7mm from the source center. |
|
کلیدواژههای انگلیسی مقاله |
Brachytherapy, high dose rate (HDR), 60Co, Monte Carlo. |
|
نویسندگان مقاله |
a مصباحی | a. mesbahi medical physics department, medical school and radiation oncology department, imam-khomeini hospital, tabriz university of medical sciences, tabriz, iran
سازمان اصلی تایید شده: دانشگاه علوم پزشکی تبریز (Tabriz university of medical sciences)
|
|
نشانی اینترنتی |
http://www.ijrr.com/browse.php?a_code=A-10-1-255&slc_lang=en&sid=en |
فایل مقاله |
فایلی برای مقاله ذخیره نشده است |
کد مقاله (doi) |
|
زبان مقاله منتشر شده |
en |
موضوعات مقاله منتشر شده |
Radiation Biology |
نوع مقاله منتشر شده |
تحقیق بدیع |
|
|
برگشت به:
صفحه اول پایگاه |
نسخه مرتبط |
نشریه مرتبط |
فهرست نشریات
|