Archives of Anesthesiology and Critical Care، جلد ۳، شماره ۴، صفحات ۴۰۰-۴۰۲

عنوان فارسی
چکیده فارسی مقاله
کلیدواژه‌های فارسی مقاله

عنوان انگلیسی Complication of Intraoperative Percutaneous Dilatational Tracheostomy without Bronchoscopy: A Case Report
چکیده انگلیسی مقاله A 72-year-old woman who was a candidate for percutaneous dilatational tracheostomy (PDT) using the Ciaglia Blue Rhino single dilator was examined. After dilation, a stoma was created on the guide wire, but the dilator could hardly pass through and pulled in the endotracheal tube with it. Bronchoscopy was therefore performed. The guide wire that was inside the Murphy eye of the endotracheal tube was then removed. So the dilator and guide wire was removed. The endotracheal tube was pulled back slightly and the tracheostomy tube was inserted once again. Fiberoptic bronchoscopy is recommended in cases of a problematic anatomy or when the PDT tube placement is suspicious.
کلیدواژه‌های انگلیسی مقاله

نویسندگان مقاله پرویز امری مله | parviz amri maleh


فاطمه دماوندی | fatemeh damavandi


فاطمه حیدرنیا | fatemeh heidarnia



نشانی اینترنتی http://aacc.tums.ac.ir/index.php/aacc/article/view/121
فایل مقاله دریافت فایل مقاله
کد مقاله (doi)
زبان مقاله منتشر شده en
موضوعات مقاله منتشر شده
نوع مقاله منتشر شده Case Report(s)
برگشت به: صفحه اول پایگاه   |   نسخه مرتبط   |   نشریه مرتبط   |   فهرست نشریات